TitleMammography images restoration by quantum noise reduction and inverse MTF filtering
FormatOn-line
Year2009
Access Date2024, May 01
Number of Files1
Size989 KiB
2. Context
AuthorRomualdo, Larissa Cristina dos Santos
AffiliationUniversidade de São Paulo - Departamento de Engenharia Elérica
EditorNonato, Luis Gustavo
Scharcanski, Jacob
e-Mail Addresslarissa.romualdo@uol.com.br
Conference NameBrazilian Symposium on Computer Graphics and Image Processing, 22 (SIBGRAPI)
Conference LocationRio de Janeiro, RJ, Brazil
Date11-14 Oct. 2009
PublisherIEEE Computer Society
Publisher CityLos Alamitos
Book TitleProceedings
Tertiary TypeFull Paper
History (UTC)2010-08-28 20:03:27 :: larissa.romualdo@uol.com.br -> administrator ::
2022-06-14 00:14:05 :: administrator -> :: 2009
3. Content and structure
Is the master or a copy?is the master
Content Stagecompleted
Transferable1
Version Typefinaldraft
Keywordsimage restoration
mammography
Anscombe transform
quantum noise
AbstractThis work proposes a new restoration method to improve mammographic images by using Anscombe Transform and Wiener Filter to quantum noise reduction. Besides, it is performed an image enhancement by using a restoration inverse filter, calculated based on the image system modulation transfer function (MTF). This pre-processing technique were used for a set of mammographic phantom images in order measure the number of microcalcifications correctly detected by a computer-aided detection (CAD) algorithm. Results showedthat the proposed method improved breast images quality by overcome the acquisition process constrains and reducing noise. The performance of the breast microcalcification CAD was improved when using the restored images set in comparison to the original one.
Arrangement 1urlib.net > SDLA > Fonds > SIBGRAPI 2009 > Mammography images restoration...
Arrangement 2urlib.net > SDLA > Fonds > Full Index > Mammography images restoration...
doc Directory Contentaccess
source Directory Contentthere are no files
agreement Directory Contentthere are no files
4. Conditions of access and use
data URLhttp://urlib.net/ibi/8JMKD3MGPBW4/35SJUJE
zipped data URLhttp://urlib.net/zip/8JMKD3MGPBW4/35SJUJE
Languageen
Target File58045.pdf
User Grouplarissa.romualdo@uol.com.br
Visibilityshown
5. Allied materials
Next Higher Units8JMKD3MGPEW34M/46SJQ2S
8JMKD3MGPEW34M/4742MCS
Citing Item Listsid.inpe.br/sibgrapi/2022/05.14.19.43 2
Host Collectionsid.inpe.br/banon/2001/03.30.15.38
6. Notes
Empty Fieldsarchivingpolicy archivist area callnumber contenttype copyholder copyright creatorhistory descriptionlevel dissemination documentstage edition electronicmailaddress group isbn issn label lineage mark mirrorrepository nextedition notes numberofvolumes orcid organization pages parameterlist parentrepositories previousedition previouslowerunit progress project readergroup readpermission resumeid rightsholder schedulinginformation secondarydate secondarykey secondarymark secondarytype serieseditor session shorttitle sponsor subject tertiarymark type url volume


Close